LY-NDIR Non-dispersive Infrared CO/CO₂/SO₂/NO/CH₄/CH /Alcohol Gas Measurement Optical Module

Module Introduction

The module is specially developed for high-quality NDIR gas analyzers. Modular design is adopted. We can also customize the functions of measuring CO, CO2 SO2, NO, CH4, HC, Alcohol Gas and other gases according to users’ requirements. The measuring range is from ppm to % level. This module has a smaller size and is convenient to be integrated into various gas analyzers.

Principle

The module adopts the principle of non-dispersive infrared absorption and consists of a wide-spectrum infrared radiation light source, a gas chamber, a detector and a signal control and processing circuit. It analyzes the gas concentration by detecting the light attenuation before and after gas absorption. The module adopts a high-performance wide-spectrum infrared radiation light source and a new type of multi-reflection infrared gas chamber designed by Laoying, achieving both high stability and low detection limit. The optical path of the gas chamber can be customized, and the chamber wall does not affect the optical process, so the optical path is more stable and the measurement results are more reliable, without frequent cleaning of the gas chamber and other maintenance.

 

Features
  • Adopting a pulse modulation infrared light source, without the need of mechanical chopping modulation;

  • Adopting-multi-channel pyroelectric detector,which provides higher sensitivity;

  • Constant temperature control ensures good stability;

  • Self-developed algorithm provides stronger anti-interference capability;

  • Measured gases, performance index and mechanical structure can be customized according to requirements.

 

Applications
  • Ambient air analysis

  • Indoor air quality monitoring

  • Analysis of waste gas from pollution sources

  • Analysis of emission from mobile pollution sources

  • Industrial gas analysis

  • Process measurement technology

  • Biogas research, etc.

  • Greenhouse gas monitoring

  • Process control

  • Building ventilation system management

  • Analysis of CO2 partial pressure and dissolved inorganic carbon

  • Atmospheric monitoring and meteorological research

Technical Indicators

 

Technical Indicators

Specifications

Measured gas

Measuring range

Measuring range (ppm)

CO

0-100%

0-150000

0-5000

0-1000

0-200

CO2

0-100%

0-200000

0-20000

0-1000

0-500

SO2

-

-

0-2500

0-1000

0-200

NO

-

0-30000

0-5000

0-1000

0-500

CH4

0-100%

0-10000

0-5000

0-1000

0-500

HC

0-10%

0-10000

0-5000

0-1000

0-500

Customizable according to user requirements

Resolution

0.01%

1ppm

0.1ppm

Indication error

≤±2%FS

Repeatability

≤2%

Zero Drift

≤±2%FS@1h

Span Drift

≤±2%FS@1h

Response time

T90 < 30 s@0.5L/min

Operating temperature

(0~40)℃

Relative humidity

≤95%RH

Calibration

Zero calibration and span calibration

Communication interface

RS232/RS485

Output signal

Concentration/Original AD value

Power supply

DC12V/24V

Power

≤ 25 W (upon power-on); ≤ 5 W (when working stably)

 



 

Explanation

The above parameters are totally up to the requirements of the relevant national standards. If product upgrades, or image does not match with real machine, please refer to the real machine.