LY-NDUV Non-Dispersive Ultraviolet O₃ Gas Measurement Optical Module
The module is specially developed for NDUV O3 gas analyzers. Modular design is adopted and range can be customized, The measurement range is from ppb to % level, Small size, high measurement accuracy, easy to integrate into the O3 gas analyzer.
The module adopts the principle of non-dispersive Ultraviolet absorption and consists of a Narrow band UV LED, a gas chamber, a detector and a signal control and processing circuit. It analyzes the gas concentration by detecting the light intensity before and after gas absorption and thus calculating the absorbance.
The module adopts a new type of multi-reflection Ultraviolet gas chamber designed by Laoying, It has the advantages of customized optical path and stable optical path, achieving both high stability and low detection limit. The module has a smaller size and is convenient to be integrated into various gas analyzers.
- UV-LED light source, low power consumption, Long service life no need to preheat
- Low detection limit, high accuracy, Wide measurement range
- Small in size and light in weight
- The mechanical structure and specifications according to customer application requirements.
- Built in reference channel, high measurement accuracy
- Ambient air analysis
- Analysis of exhaust gas from stationary sources
- Analysis of emission from mobile pollution Sources
- Industrial gas analysis
Technical Indicators |
Specifications |
||
Measured gas |
O3 |
||
Measuring range |
(0~1000)ppb |
0~1000ppm |
0~10% |
Indication error |
≤±3% FS |
||
Repeatability |
≤1% |
||
Calibration |
Zero calibration and span calibration |
||
Response time |
T90≤20s@0.5L/min |
||
Operating temperature |
(5~40)℃ |
||
Relative humidity |
≤85%RH |
||
Communication interface |
RS232/RS485 |
||
Output signal |
Concentration/Original AD value |
||
Power supply |
DC 12V/24V |
||
Power |
<10W |
||
Dimensions |
130mm×60mm×45mm |
||
Weight |
<0.3Kg |